CDR DrinkLab is a quick, reliable and easy analysis system for the quality and process control of the beverage sector

Determination of Acetic Acid in Molasses

Acetic acid is a particularly important parameter in molasses because it is one of the most critical acids affecting the quality of the final product. Acetic acid has a strong impact on yeast activity. Even at relatively low concentrations it can be toxic: it enters the cell in its undissociated form and causes metabolic stress, reducing fermentation rate and alcohol yield. If its content is high, yeasts may slow down drastically or stop completely.

From a microbiological perspective, the presence of acetic acid is often an indicator of contamination or degradation processes that occurred before fermentation. It may originate from the activity of acetic or lactic acid bacteria, from improper molasses storage, or from spontaneous fermentations. For this reason, a high acetic acid value indicates a raw material that is already “compromised” or at risk.

Method

Test type: End Point.

Testing time: 6 minutes.

Analysis sessions with multiple samples are possible up to a maximum of 4 tests for CDR DrinkLab®.

...

Sample treatment

Take 900 µL of decolorizing solution *300755 and transfer it to a centrifuge tube (shake the decolorizing solution before use). Add 100 µL of the sample to be analyzed. Shake for 5 minutes and centrifuge (code 222061) for 10 minutes at 12,000 rpm.

Principle of the test

Through an enzymatic reaction, acetic acid is converted into piruvic acid, which in turn reacts with NADH producing NAD+ and D-Lactic acid. The reduction of absorbance, measured at 366 nm (end-point), of the NADH solution is proportional to the quantity of acetic acid present in the sample.

...

Calibration curve

The calibration curve of the CDR DrinkLab® system confirmed excellent linearity across the entire measurement range.

...

Reagent test Kits

Code *300359: 100 test
Code *300360: 10 test
The reagents have a shelf-life of 6 months

Measuring range

AnalysesMeasuring rangeResolutionRepeatability
Acetic acid [0.5-2.5] g/L0.5 - 2.5 g/L Acetic acid0.1 g/L Acetic acid0.1 g/L Acetic acid

The perfect system for you

image

CDR DrinkLab®

  • Complete analysis panel, supplied already configured

  • Up to 16 determinations simultaneously

  • Possibility of carrying out analyses of the same sample

  • Integrated printer

  • Full connections (LAN - USB - Bluetooth barcode/QR code reader)