CDR FoodLab® is the chemical analyzer for quality and process control that allows you to determine a wide range of parameters of many foods, animal feeds and pet foods.

Determination of Ammonia in Silage and Animal Feed

The determination of ammonia in animal feed is a parameter of major importance, as it provides essential information on protein quality, preservation status, and feed safety. This aspect is particularly relevant in moist feeds, silages, and fermented feeds. Ammonia is mainly formed as a result of protein and amino acid degradation processes, caused by the activity of undesirable microorganisms or by suboptimal storage conditions. A high ammonia content is therefore indicative of advanced proteolysis and signals abnormal fermentation or feed deterioration. This leads to a reduction in nutritional value, as protein nitrogen is converted into forms that are less available to the animal.

Method

Test type: End Point.

Testing time: 8 minutes.

Analysis sessions with multiple samples are possible up to a maximum of 16 tests for CDR FoodLab® and 3 tests for CDR FoodLab® Junior.

...

Sample treatment

Weigh 10 g of sample and transfer it into a Stomacher filter bag. Add 100 mL of distilled water and mix in the Stomacher for at least 3 minutes. Filter using a filtration kit if necessary (code 300210). Collect a few milliliters of the filtrate and allow it to settle in a test tube for a few minutes. Take 50 µL of the supernatant for analysis. Opalescence does not interfere with the reaction.

Principle of the test

The ammonium ions react with a phenolic derivative and form a blue-green colored complex whose intensity, measured at 700 nm, is directly proportional to the concentration of ammonia in the sample.

...

Calibration curve

The calibration curve of the CDR FoodLab® system confirmed excellent linearity across the entire measurement range.

...

Reagent test Kits

Code *300050: 100 test
Code *300054: 10 test
The reagents have a shelf-life of 12 months

Measuring range

AnalysesMeasuring rangeResolutionRepeatability
NH3 silage10 - 600 ppm1 ppm7 ppm

The analyzers for oils and fats quality control

image

CDR FoodLab®

  • Complete analysis panel, supplied already configured

  • Up to 16 determinations simultaneously

  • Possibility of carrying out analyses of the same sample

  • Integrated printer

  • Full connections (LAN - USB - Bluetooth barcode/QR code reader)

image

CDR FoodLab® Jr

  • Partial analysis panel, customisable configuration

  • Up to 3 determinations simultaneously

  • Wireless connection to external printer

  • USB connections