Sugars analysis on Potatoes

Glucose and fructose are the most important sugars to analyze in tubers both as a qualitative parameter and as a process control. These sugars can be precursors of acrylamide in frying products; knowing their value on potatoes, before being processed, allows to avoid problems during cooking. The glucose and fructose content can be determined together (kit codes *300380 and *300384) or separately (kit codes *300335 and *300355).

Method

Test type: End Point.

Testing time: 4 minutes.

Analysis sessions with multiple samples are possible up to a maximum of 16 tests for CDR FoodLab® and 3 tests for CDR FoodLab® Junior.

Sample treatment

Mix the sample to obtain a homogeneous pulp. Carry out a 1:1 dilution of the sample with distilled water and add 70% perchloric acid in these proportions: 100g of potato + 100g of water + 0.6 ml of perchloric acid. Centrifuge for 5 minutes at 12000 rpm. Collect the supernatant for analysis.

Principle of the test

The fructose and glucose present in the sample are determined enzymatically. The Redox reaction is read at 366 nm (end-point) and the value is directly proportional to the concentration of glucose and fructose in the sample.

...

Calibration curve

The calibration of the test was carried out by analyzing standard solutions of glucose and fructose at various concentrations.

...

Reagent test Kits

Code *300335: 100 test
Code *300355: 10 test
The reagents have a shelf-life of 15 months
Code *300380: 100 test
Code *300384: 10 test
The reagents have a shelf-life of 15 months

Measuring range

AnalysesMeasuring rangeResolutionRepeatability
Glucose fructose [5-2500] mg/Kg5 - 2500 mg/Kg Glucose1 mg/Kg Glucose28 mg/Kg Glucose
Sugars(glu,fru)[5-2500]mg/Kg5 - 2500 mg/Kg1 mg/Kg28 mg/Kg

The analyzers for oils and fats quality control

image

CDR FoodLab®

  • Complete analysis panel, supplied already configured

  • Up to 16 determinations simultaneously

  • Possibility of carrying out analyses of the same sample

  • Integrated printer

  • Full connections (LAN - USB - Bluetooth barcode/QR code reader)

image

CDR FoodLab® Jr

  • Partial analysis panel, customisable configuration

  • Up to 3 determinations simultaneously

  • Wireless connection to external printer

  • USB connections